Highlights & Features
Safety Approvals to IEC 60601-1 3rd ed. & IEC 60950-1
Risk management report available
Low touch current (<0.1mA Normal & ,0.3mA single fault)
Over-Voltage/Load/Temperature & Short Circuit protections
1 Million Hours MTBF
2 x MOPP (means of patient protection)
3 years warranty
Nominal Output Voltage24V
Output Current3.75A
Output Power90W
Line Regulation±0.5% max.
Load Regulation±5% max.
Start-up Time3000ms
Hold-up Time6ms @ 115Vac&230Vac
Dynamic Response± 5% @ 50-100% load (Overshoot & Undershoot O/P Voltage)
Ripple & Noise107.2mV pk-pk @ Full load (Ref.Datasheet)
Nominal Input Voltage100-240Vac
Input Voltage Range90-264 Vac
Nominal Input Frequency50-60Hz
Input Frequency Range47-63Hz
Input Current2A @ 115Vac, 1A @ 230Vac
Efficiency at 100% load87.61% @ Full load
Case Chassis / Case CoverPC
Dimensions (W x L x H)75 x 150 x 40 mm
Unit Weight500 g
Input TerminalC8 socket
Output TerminalBarrel type
ENVIRONMENT / PROTECTIONS / RELIABILITY / SAFETY / EMC
Operating Ambient Temperature0°C to +40°C
Storage Ambient Temperature-40°C to +85°C
Overvoltage Protection150%, Latch Mode (max.)
Overload / Overcurrent Protection150% of rated load current (max.), Hiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature ProtectionLatch Mode or Auto-Recovery
Short Circuit ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Degree of ProtectionIP22
Protection Against ShockClass II
MTBF1 Million Hrs based on Telecordia SR-332
Galvanic IsolationInput to Output 4000 Vac
Electrical SafetyIEC60601-1: (Ed.3,2005), EN0601-1:2006, CAN/CSA-C22.2 No. 60601-1:08, ANSI/AAMI ES60601-1: (Ed.3,2005)
ITEIEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
CEMDD Directive 93/42/EEC