Highlights & Features
Compliant to IEC 60601-1 3rd edition
IT and medical safety approvals
Low earth leakage current (<0.1mA)
Overload, over voltage, over temperature and short circuit protections
Risk management report available
Energy Star Compliance (Level V)
2 x MOPP(means of patient protection)
Nominal Output Voltage19V
Output Current3.2A
Output Power60W
Line Regulation± 0.5%
Load Regulation± 5%
Start-up Time3000ms @ 115Vac
Hold-up Time10ms @ 115Vac
Dynamic Response± 5% @ 50-100% load
Ripple & Noise61mV pk-pk @ Full load (typ.)
Nominal Input Voltage100-250Vac
Input Voltage Range90-275 Vac
Nominal Input Frequency50-60Hz
Input Frequency Range47-63Hz
Input Current1.5A @ 115Vac,
0.75A @ 230Vac
Efficiency at 100% load88% (typ.)
Max. Inrush Current50A @ 115Vac,
100A @ 230Vac (typ.)
Leakage Current0.1mA @ 275Vac NC,
0.3mA @ 275Vac SFC
Nominal Current1.5A at 90Vac -0.75A at 264Vac
Case Chassis / Case CoverPC
Dimensions (W x L x H)62 x 135 x 34.1 mm
Unit Weight0.36 kg
Input TerminalC14 socket
Output Terminal4 Pin Din
ENVIRONMENT / PROTECTIONS / RELIABILITY / SAFETY / EMC
Operating Ambient Temperature0°C to +40°C
Storage Ambient Temperature-40°C to +85°C
Operating Humidity10-95% RH (Non-Condensing)
Shock Test (Non-Operating)50G, 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)5-500Hz, 2.09Grms, 20 minute for each three axis
Overvoltage Protection150%, Latch Mode
Overload / Overcurrent Protection180% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Degree of ProtectionIP40
Protection Against ShockClass I
MTBF1218 kHrs based on Telecordia SR-332
Galvanic IsolationInput to Output: 4000Vac
Input to Ground: 1500Vac
Output to Ground: 500Vac
EmissionsEN55011, EN55022, FCC Title 47: Class B
Voltage Fluctuation and FlickerIEC61000-3-3
ITEIEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
CEMDD Directive 93/42/EEC
Radiated ImmunityEN61000-4-3, 3V/m Perf Criteria A 80MHz-1GHz, 3V/M with 1kHz tone / 80% modulation
EFTEN61000-4-4, Level 3 Criteria A 2kV
SurgeEN61000-4-5, Level 3 Criteria A, Common Mode: 2kV, Differential Mode: 1kV
Conducted ImmunityIEC61000-4-6, Level 2 Criteria A 150kHz-80MHz, 3Vrms
Voltage DipsIEC61000-4-11, 30% 10ms Criteria A ; 60% 100ms and 100% 5000ms Criteria B