Highlights & Features
Compliant to IEC 60601-1 3rd edition
IT and medical safety approvals
Low earth leakage current (<0.1mA)
Overload, over voltage, over temperature and short circuit protections
Risk management report available
Energy Star Compliance (Level V)
2 x MOPP(means of patient protection)
Dimensions (W x L x H): 85 x 170 x 40 mm
Unit Weight: 1.1 Kg
Nominal Output Voltage12V
Output Current10.0A
Output Power120W
Line Regulation± 0.5%
Load Regulation± 5%
Start-up Time3000ms @ 115Vac
Hold-up Time20ms @ 115Vac
Dynamic Response± 5% @ 50-100% load
Ripple & Noise79mV @ Full load (typ.)
Nominal Input Voltage100-240Vac
Input Voltage Range90-264 Vac
Nominal Input Frequency50-60Hz
Input Frequency Range47-63Hz
Input Current4A @ 115Vac,
2A @ 230Vac
Efficiency at 100% load88.1% (typ.)
Max. Inrush Current50A @ 115Vac,
100A @ 230Vac (typ.)
Leakage Current0.1mA @ 275Vac NC,
0.3mA @ 275Vac SFC
Case Chassis / Case CoverPC
Dimensions (W x L x H)85 x 170 x 40 mm
Unit Weight1.1 Kg
Input TerminalC14 socket
Output Terminal4 Pin Din
ENVIRONMENT / PROTECTIONS / RELIABILITY / SAFETY / EMC
Operating Ambient Temperature0°C to +40°C
Storage Ambient Temperature-40°C to +85°C
Operating Humidity10-95% RH (Non-Condensing)
Shock Test (Non-Operating)50G, 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)5-500Hz, 2.09Grms, 20 minute for each three axis
Overvoltage Protection130% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery)
Overload / Overcurrent ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Degree of ProtectionIP40
Protection Against ShockClass II with functional earth
MTBF1185 kHrs based on Telecordia SR-332
Galvanic IsolationInput to Output: 4000 Vac
EmissionsEN55011/22/FCC conducted / radiated class B
Voltage Fluctuation and FlickerEN61000-3-3
ITEIEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
CEMDD Directive 93/42/EEC
Radiated ImmunityEN61000-4-3, 3V/m Perf Criteria A 80MHz-1GHz, 3V/M with 1kHz tone / 80% modulation
EFTEN61000-4-4, Level 3 Criteria A 2kV
SurgeEN61000-4-5, Level 3 Criteria A, Common Mode: 2kV, Differential Mode: 1kV
Conducted ImmunityEN61000-4-6, 3Vrms Perf Criteria A
Voltage DipsIEC61000-4-11, 30% 10ms Criteria A ; 60% 100ms and 100% 5000ms Criteria B