Highlights & Features
Safety Approvals to IEC 60601-1 3rd ed. & IEC 60950-1
Risk management report available
Low touch current (<0.1mA Normal & ,0.3mA single fault)
Over-Voltage/Load/Temperature & Short Circuit protections
1 Million Hours MTBF – Small Size and Low Weight
2 x MOPP (means of patient protection)
3 years warranty
Nominal Output Voltage5V
Output Current1.0A
Output Power5W
Line Regulation±5% (max.)
Load Regulation±10% (max.)
Start-up Time3000ms @ 115Vac (max.)
Hold-up Time5ms @ 115Vac (min.)
Dynamic Response±10% @ 50-100% load (Overshoot & Undershoot O/P Voltage)
Nominal Input Voltage100-240Vac
Input Voltage Range90-264 Vac
Nominal Input Frequency50-60Hz
Input Frequency Range47-63Hz
Input Current0.2A @ 115Vac ; 0.1A @ 230Vac
Efficiency at 100% load71% (typ.)
Case Chassis / Case CoverPC / PC
Dimensions (W x L x H)39.5 x 56.5 x 28 mm
Unit Weight100 grams
Input TerminalWall mount - US & EU Type
Output TerminalBarrel type
ENVIRONMENT / PROTECTIONS / RELIABILITY / SAFETY / EMC
Operating Ambient Temperature0°C to +40°C
Storage Ambient Temperature-40°C to +85°C
Operating Humidity5-95% RH (Non-Condensing)
Shock Test (Non-Operating)50G, 11ms, 3 shocks for each direction
Overvoltage Protection150% of rated output voltage
Overload / Overcurrent Protection130% of rated load current (Non-Latching, Auto-Recovery)
Over Temperature ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Degree of ProtectionIP22
Protection Against ShockClass II
MTBF1 Million Hours based on Telecordia SR-332