Highlights & Features
Compliant to IEC 60601-1 3rd edition
IT and medical safety approvals
Low earth leakage current (<0.1mA)
Overload, over voltage, over temperature and short circuit protections
Risk management report available
Nominal Output Voltage24V
Output Current16.67A
Output Power400W
Line Regulation±0.5%
Hold-up Time12ms @ 115Vac
Ripple & Noise121.6mV pk-pk @ Full load (typ.)
Nominal Input Voltage100-240Vac
Input Voltage Range90-264 Vac
Nominal Input Frequency50-60Hz
Input Frequency Range47-63Hz
Input Current5.5A @ 90Vac,
2.7A @ 264Vac
Input Power (Standby Mode)1.2W
Efficiency at 100% load91.2% (typ.)
Max. Inrush Current60A @ 115Vac,
60A @ 230Vac (typ.)
Leakage Current0.1mA @ 240Vac Normal Condition,
0.3mA @ 264Vac Single Fault Condition
Nominal Current5.5A at 90Vac -2.7A at 264Vac
Case Chassis / Case CoverAL1100F(+Anode)
Dimensions (W x L x H)97 x 198 x 41.5 mm
Unit Weight0.91 kg
Input TerminalDECA T33-BM1103A301
Output TerminalDECA T33-BM1104A301
ENVIRONMENT / PROTECTIONS / RELIABILITY / SAFETY / EMC
Operating Ambient Temperature-10°C to +70°C
Storage Ambient Temperature-40°C to +85°C
Power De-rating-10°C to +50°C 100% load
+50°C to +70°C de-rate power by 2.5% / °C
Operating Humidity10-95% RH (Non-Condensing)
Shock Test (Non-Operating)50G 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)5-500Hz, 2.09Grms, 20minutes for each three axis
Overvoltage Protection150%, Latch Mode
Overload / Overcurrent Protection150% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Protection Against ShockClass I with PE connection
MTBF1998 kHrs based on Telcordia SR-332
EmissionsEN55011, EN55022, FCC Title 47: Class B
ITEIEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
CEMDD Directive 93/42/EEC
EFTEN61000-4-4, Level 3 (2kV) Perf Criteria A
SurgeEN61000-4-5, Level 3 Criteria A Common Mode: 2kV Differential Mode: 1kV
Conducted ImmunityEN61000-4-6, Level 2 Criteria A 150kHz-80MHz, 3Vrms
Magnetic FieldEN61000-4-8, Criteria A Magnetic field strength 3A/Meter
Voltage DipsConform to EN61000-4-11, 30% 10ms Criteria A,60% 100ms and 100% 5000ms Criteria B