Highlights & Features
Compliant to IEC 60601-1 3rd edition
IT and medical safety approvals
Low earth leakage Current (<0.1 mA)
Overload, over voltage, over temperature and short circuit protections
Risk management report available
Nominal Output Voltage24V
Output Current1.7A
Output Power40W
Line Regulation± 0.5%
Load Regulation± 1%
Start-up Time3000ms @ 115Vac
Hold-up Time15ms @ 115Vac
Dynamic Response± 5% @ 50-100% load
Ripple & Noise34 mV pk-pk @ Full load (typ.)
Nominal Input Voltage100-240Vac
Input Voltage Range90-264 Vac
Nominal Input Frequency50-60Hz
Input Frequency Range47-63Hz
Input Current0.85A @ 115V
Efficiency at 100% load86.3% (typ.)
Max. Inrush Current< 30A @ 115Vac,
< 60A @ 230Vac (typ.)
Leakage Current0.1mA @ 264Vac Normal Condition
0.3mA @ 264Vac Single Fault Condition
Nominal Current1.0A @ 90Vac
< 0.5A @ 264Vac
Dimensions (W x L x H)50.8 x 76.2 x 22.9 mm
Unit Weight0.08 kg
Input TerminalJST 2P
Output TerminalJST 4P
ENVIRONMENT / PROTECTIONS / RELIABILITY / SAFETY / EMC
Operating Ambient Temperature0°C to +40°C
Storage Ambient Temperature-40°C to +85°C
Operating Humidity10-95% RH (Non-Condensing)
Shock Test (Non-Operating)50G 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)5-500Hz, 2.09Grms, 20 minutes for each three axis
Overvoltage Protection150%, Latch Mode
Overload / Overcurrent Protection150%~210% of rated load current, Hiccup Mode, Non-Latching, Auto-Recovery)
Over Temperature ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Reverse Polarity ProtectionClass I with PE* connection
MTBF3257 kHrs based on Telecordia SR-332
EmissionsEN55011/22/FCC conducted / radiated class B
Voltage Fluctuation and FlickerEN61000-3-3
ESD ImmunityEN61000-4-2,6kV contact / 8kV aair class A
Radiated ImmunityEN61000-4-3, 3V/m Perf Criteria A
EFTEN61000-4-4, Level 3 (2kV) Perf Criteria A
SurgeEN61000-4-5, DM± 1kV /CM ± 2kV Perf Criteria A
Conducted ImmunityEN61000-4-6, 3Vrms Perf Criteria A
Magnetic FieldEN61000-4-8, 3A/m Perf Criteria A
Voltage DipsConform to EN61000-4-11, 30% 10ms Criteria A, 60% 100ms and 100% 5000ms Criteria B