Highlights & Features
Compliant to IEC 60601-1 3rd edition
IT and medical safety approvals
Low touch leakage current (<0.1mA)
Overload, over voltage, over temperature and short circuit protections
Risk management report available
Nominal Output Voltage12V
Output Current8.33A
Output Power100W
Line Regulation±0.5%
Load Regulation±1%
Start-up Time3000ms @ 115Vac
Hold-up Time10ms @ 115Vac
Dynamic Response± 3% @ 50-100% load
Ripple & Noise76mV pk-pk @ Full load
Nominal Input Voltage100-240Vac
Input Voltage Range90-264 Vac
Nominal Input Frequency50-60Hz
Input Frequency Range47-63Hz
Input Current2.5A @ 100Vac, 1.25A @ 240Vac
Efficiency at 100% load87%
Max. Inrush Current30A @ 115Vac, 60A @ 230Vac (typ.)
Leakage Current0.1mA @ 264Vac NC, 0.3mA @ 264Vac SFC
Dimensions (W x L x H)50.8 x 101.6 x 31.8 mm
Unit Weight0.15 kg
Input TerminalJST 2P
Output TerminalJST 4P
ENVIRONMENT / PROTECTIONS / RELIABILITY / SAFETY / EMC
Operating Ambient Temperature-10°C to +70°C
Storage Ambient Temperature-40°C to +85°C
Power De-rating-10°C to +50°C 100% load
50°C to 70°C de-rate power by 2.5%
Operating Humidity10-95% RH (Non-Condensing)
Shock Test (Non-Operating)50G, 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)5-500Hz, 2.09Grms, 20 minute for each three axis
Overvoltage Protection150%, Latch Mode (max.)
Overload / Overcurrent Protection170% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery) (max.)
Over Temperature ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Protection Against ShockClass I with Primary Earth connection
MTBF983 kHrs based on Telecordia SR-332
Safety ApprovalsIEC60601-1: (Ed.3,2005), EN0601-1:2006, CAN/CSAC22.2 No. 60601-1:08, ANSI/AAMI ES60601-1: (Ed.3,2005)
Galvanic IsolationInput to Output: 4000 Vac
Input to Ground: 1500 Vac
Output to Ground: 500 Vac
EmissionsEN55011, EN55022, FCC Title 47: Class B
Voltage Fluctuation and FlickerIEC61000-3-3