5.1V/ 12V/ -15V/ 15V 100W / MDS-100AP401 B
Highlights & Features
Safety Approvals to IEC 60601-1 3rd ed. & IEC 60950-1
Risk management report available
Low touch current (<0.1mA Normal & ,0.3mA single fault)
Over-Voltage/Load/Temperature & Short Circuit protections
1 Million Hours MTBF
2 x MOPP (means of patient protection)
3 years warranty
Nominal Output Voltage5.1V/ 12V/ -15V/ 15V
Output Current8.0A / 3.0A / 0.8A / 0.8A
Output Power100W
Line Regulation±0.5%
Load RegulationV1=±2%, V2=±1%, V2=±3%, V4=±3%
Start-up Time3000ms max @ 115Vac
Hold-up Time5ms min @ 115Vac
Dynamic Response± 5% @ 50-100% load
Nominal Input Voltage100-240Vac
Input Voltage Range90-264 Vac
Nominal Input Frequency50-60Hz
Input Frequency Range47-63Hz
Input Current2.5A @ 115Vac,1.3A@230Vac
Efficiency at 100% load79.34%
Max. Inrush Current30A @ 115Vac, 60A @ 230Vac
Leakage Current0.1mA @ 264Vac
Case Chassis / Case CoverNA
Dimensions (W x L x H)76.2 x 127 x 30.5 mm
Unit Weight0.25 kg
Input TerminalMolex 2P
Output TerminalMolex 12P
ENVIRONMENT / PROTECTIONS / RELIABILITY / SAFETY / EMC
Operating Ambient Temperature-10°C to +70°C
Storage Ambient Temperature-40°C to +85°C
Operating Humidity5-95% RH (Non-Condensing)
Shock Test (Non-Operating)50G, 11ms, 3 shocks for each direction
Overvoltage Protection150%, Latch Mode
Overload / Overcurrent Protection150%~200% of rated load current, Hiccup Mode,
(Non-Latching, Auto-Recovery)
Over Temperature ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit ProtectionHiccup Mode, (Non-Latching, Auto-Recovery)
Protection Against ShockClass I with PE connection
MTBF1.4 Million Hours based on Telecordia SR-332
EmissionsEN55011, EN55022, FCC Title 47: Class B
Harmonic Current EmissionsIEC61000-3-2
Electrical SafetyIEC60601-1:2005, EN0601-1:2006, CAN/CSA-C22.2 No. 60601-1:08,
ANSI/AAMI ES60601-1:2005
ITEIEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
CEMDD Directive 93/42/EEC